We have developed a simultaneous quantification method for prostaglandin (PG) E(2), PGD(2), PGF(2 alpha), 8-epi-PGF(2 alpha), 6-keto-PGF(1 alpha) and thromboxane (TX) B(2). Using [3,3,4,4-(2)H(4)]PGE(2), [3,3,4,4-(2)H(4)]PGD(2), [3,3,4,4-(2)H(4)]8-epi-PGF(2 alpha), [3,3,4,4-(2)H(4)]PGF(2 alpha), [3,3,4,4-(2)H(4)]6-keto-PGF(1 alpha) and [18,18,19,19-(2)H(4)]TXB(2) as internal standards (I.S.), the eicosanoids and their I.S. were simultaneously extracted by solid-phase extraction from cell-cultured medium, derivatized to methyl ester/methoxim/tert.-butyldimethylsilyl ether derivatives and analyzed using gas chromatography-mass spectrometry in the selected ion monitoring mode. The accuracy for the added eicosanoids ranged from 92 to 113%, and coefficients of variation ranged from 0.1 to 12.2%. Increased eicosanoids in RAW264.7 and U937 cells stimulated by lipopolysaccharide were suppressed by NS-398 and indometacin. This simultaneous quantification method can be applied routinely for assaying eicosanoids in vitro.
774(2):205-14.